Milwaukee Mi170 Auto-Ranging EC/TDS/NaCl/Temperature Laboratory Bench Meter

Manual

Milwaukee Mi170 Included Accessories

1 - MA813D/1 EC/TDS/NaCl/Temperature Probe

1 - MA9315 Electrode Holder

1 - M10030 12880 µS/cm calibration solution

1 - M10031 1413 µS/cm calibration solution

1 - Mi5200 Application Software

1 - MA9350 RS232 connection cable with 2 meters cable

1 - 12 VDC Adapter

1 - Instruction manual

- Measures 4 different parameters - EC, TDS (Total Dissolved Solids), % of NaCl & temperature
- Auto-Ranging for EC & TDS Measurements
- Automatic temperature compensation
- 4-ring probe
- PC Compatible
The new Milwaukee Mi170 measures 4 different parameters - EC, TDS (Total Dissolved Solids), percentage of NaCl and temperature in a variety of ranges. The auto-ranging feature for EC and TDS measurements automatically sets the resolution suitable to the tested sample. All measurements can be temperature compensated at 20 or 25°C and the compensation coefficient is selectable by the user.

The automatic temperature compensation can also be disabled for measuring the actual conductivity value. The stability indicator on the LCD ensures accuracy.

Conductivity readings are performed with the 4-ring probe supplied with the meter. The GLP feature allows users to store and recall data on system status. PC compatible through an RS232 or USB port.

Note: We ship to the USA and Canada only.

Specifications:

EC Range

0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm actual Conductivity

TDS Range

0.00 to 14.99 mg/l (ppm); 15.0 to 149.9 mg/l (ppm); 150 to 1499 mg/l (ppm); 1.5 to 14.99 g/l (ppt); 15.0 to 100.0 g/l (ppt); up to 400.0 g/l actual TDS (with 0.80 factor)

NaCl Range

0.0 to 400.0%

Temp Range

-20.0 to 120.0°C / -4.0 to 248.0°F

EC Resolution

0.01 µS/cm; 0.1 µS/cm; 1.0 µS/cm; 0.01 mS/cm; 0.1 mS/cm

TDS Resolution

0.01 mg/l; 0.1 mg/l; 1.0 mg/l; 0.01 g/l; 0.1 g/l

NaCl Resolution

0.1%

Temp Resolution

0.1°C / 0.1°F

EC Accuracy

±1% of reading ±(0.05 µS/cm or 1 digit)

TDS Accuracy

±1% of reading ±(0.03 mg/l or 1 digit)

NaCl Accuracy

±1% of reading

Temp Accuracy

± 0.4°C / ± 0.8°F

EC Calibration

1 point slope calibration with 6 memorized solutions (84.0 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm)

NaCl Calibration

1 point, with MA9066 calibration solution

Temp Calibration

2 points, 0 to 50°C / 32 to 12 °F

Temp Compensation

automatic or manual, from -20.0 to 120.0°C / -4.0 to 248.0°F

Temp Coefficient

selectable from 0.00 to 6.00%/°C (EC and TDS only)

Probe

MA813D/1 4-ring probe with built-in temperature sensor (included)

TDS Factor

0.40 to 0.80 (default value is 0.50)

Log on Demand

up to 50 samples on each range (EC, TDS, NaCl)

GLP

last EC, NaCl calibration data

PC Interface

RS232 / USB Opto-isolated

Environment

0 to 50°C / 32 to 122°F; max RH 95%

Power supply

12 VDC power adapter (included)